Thermometer in cryogenic design - Innovative design for low-temperature measurements
Applications
- Extremely-low temperature measurements
- Applications with liquid hydrogen (LH2)
- Cryogenic tanks
- Cryogenic pipelines
- Hydrogen liquefaction plants
Special features
- Innovative design with high accuracy
- Usable with all relevant thermometers
- Approved for applications in hazardous areas
Description
Due to the global use of hydrogen, extremely low-temperature measurement during transport and storage is gaining an ever-greater relevance. With the cryo design, WIKA provides a suitable option for all relevant temperature measuring instruments for this demanding application.
In laboratory trials, resistance thermometers (Pt1000) and thermocouples (type E) were tested for their suitability in cryogenic applications. The special design of the cryogenic thermometer features a high accuracy of ±3 Kelvin at -253 °C [-423 °F] as well as high reproducibility.
The test data from laboratory investigations served as the basis for calculating new polynomials for Pt1000 resistance thermometers in the range of -258 ... -200 °C [-432 ... -328 °F] which are used in the configuration of WIKA transmitters.
What is the permissible minimum bending radius for an MI cable?
VDI/VDE 3511 Sheet 2 recommends a radius of curvature R of ≥ 5 x D (D=external diameter of the MI cable), some manufacturers of MI cable even give ≥ 3 x D as the minimum bending radius.
What are 2-, 3- and 4-wire circuits?
They describe the number of wires with which the measuring resistor (e.g. a Pt100) is connected. While with the simplest 2-wire connection, the lead resistance can falsify the measuring result, this negative influence can be compensated within the ...